Gold exploration with portable X-ray diffraction (pXRD)

In the field or at the core shed, geologists are often constrained to what minerals they can see with their hand lens. Due to the fine-grained nature of alteration that often accompanies gold mineralization, visual assessments made by the geologist are often subjective and, frequently, wrong. Samples are often sent off site to a laboratory for petrographic analysis. This procedure is time-intensive and expensive and sometimes fails to distinguish between minerals that have vastly different origins (e.g. formed by ground water or mineralizing fluids). Further, differentiating similar minerals, such as clays, carbonates, and fine-grained sulfides, is extremely difficult under the microscope.

Gold exploration with portable X-ray diffraction (pXRD)

Evident pXRD provides geologists with reliable, qualitative, and quantitative mineralogical data in near real time in the field, at the drill-rig, or in the core-shed. The information that pXRD provides is objective, rather than subjective, and would take weeks or months to obtain using traditional methods.

Using an Evident pXRD instrument, geologists can quickly make important decisions, such as:

  • Whether to abandon or extend the drill hole
  • Where to place the next drill hole
  • Whether to continue mapping in the area

Aside from alteration vectoring, quantitative mineralogy provides important geo-metallurgical and mineral processing information for future mine design.

Evident pXRD instruments use patented technology developed for the Mars Science Laboratories (MSL) mission. They are fitted with a charge-coupled device (CCD) to simultaneously collect X-ray diffraction and qualitative X-ray fluorescence data.

The benefits of Evident’ innovative pXRD analyzers include:

  • Small sample: requires only 15 mg of material
  • Easy sample preparation: does not require a skilled technician
  • Fast acquisition time: obtain results in a few minutes
  • Portability: battery-operated, rugged design with no moving parts
  • Stand-alone instrument: no need for water cooling or a large, external power source
  • No ongoing servicing requirements: enables XRD to be performed regularly without downtime

Gold exploration with portable X-ray diffraction (pXRD)

Figure 1. Alteration model through an epithermal gold deposit. Alteration facies were developed and mapped using quantitative mineralogy data derived from an Evident pXRD instrument.

Burkett et al. (2015).*

Evident XRD — Data quality and data collection time

Evident pXRD analyzers provide the convenience of onsite quantitative mineralogy, without the need for complicated sample preparation procedures. To illustrate the quantitative capabilities of Evident pXRD products, comparisons have been made with a large, 4 kW, conventional lab-based XRD instrument (Figure 2). The results show strong agreement between the lab-based XRD unit and the Evident pXRD system.

Gold exploration with portable X-ray diffraction (pXRD)

Figure 2. Excellent correlation between mineral percentages deduced from XRD traces from a lab-based, 4 kW XRD instrument (x-axis) and the Evident pXRD (y-axis) with 20 minute runtimes. Burkett et al. 2015.*

Data collection time was also studied to assess the performance of the Evident pXRD analyzers at varying testing times as well as comparison with lab-based XRD. Diffractograms of one sample using different testing times are shown in Figure 3. Figure 4 and Table 1 summarize the quantitative results at different data collection times and clearly shows consistent results between the two instruments (even at a runtime of 5 minutes).

Gold exploration with portable X-ray diffraction (pXRD)

Figure 3. XRD traces obtained from a lab-based, 4 kW XRD instrument and the Evident pXRD at varying collection times. Burkett et al. 2015.*

Results of quantitative XRD for a 4 kW, lab-based instrument and the Evident pXRD at varying collection times

Mineral Lab-based XRD pXRD - 40 mins pXRD - 20 mins pXRD - 10 mins pXRD - 5 mins
Pyrite 0.8 0.8 0.1 0.0 0.0
Calcite 6.1 5.4 7.1 8.1 8.4
Adularia 12.7 12.0 13.7 13.1 13.0
Albite 44.6 43.4 46.6 45.6 44.4
Muscovite 4.2 6.9 6.3 6.0 7.3
Chlorite 18.1 18.1 14.2 15.5 15.5
Quartz 13.5 13.5 12.1 11.6 11.4

Gold exploration with portable X-ray diffraction (pXRD)

Figure 4. Comparisons between mineral percentages deduced from the XRD traces in Figure 3. Note the excellent relationship between the Evident pXRD and lab-based instrument, even at short runtimes.

* D Burkett, I Graham, L Spencer, P Lennox, D Cohen, H Zwingmann, F Lau, B Kelly and D Cendon. (2015). “The Kulumadau Epithermal Breccia-hosted Gold Deposit, Woodlark Island, Papua New Guinea.” Pacific Rim Congress 2015, Hong Kong, China. The Australian Institute of Mining and Metallurgy, pages 1–8.

Products Used for This Application

The TERRA™ II XRD analyzer provides portable quantitative mineralogy. Featuring a battery life up to six hours and a rugged, weatherproof case, the system is built for fast in-field analysis of major and minor components.

BTX™ III Benchtop XRD Analyzer

The BTX™ III XRD analyzer offers reliable quantitative mineralogy of major and minor components in a compact, benchtop design. Powerful software is paired with an improved X-ray detector for increased speed and sensitivity.