Application Notes

Contact Us

Inspecting Train Wheels Using Phased Array

Inspecting Train Wheels Using Phased Array

Background

Stringent regulations are imposed on the inspection of train wheels in both the manufacturing and service sectors. To meet and exceed these strict criteria, Evident provides powerful and flexible ultrasonic instruments, dedicated software, and turnkey systems to inspect cast or forged wheels during manufacturing or while they are in service after reprofiling.

Typical Inspection Procedure

Post-Manufacturing Inspection

When cast or forged wheels are manufactured, the rim is inspected using ultrasonic phased array technology. The rest of the wheel is inspected using magnetic particle testing.

The rim is inspected by placing one linear phased array probe on the tread and a second on the rim’s side (Figure 1).

During this process, the hub can also be inspected using a third linear phased array probe. It is also possible to inspect the hub from both sides (Figure 2).

The rim is inspected using two linear phased array probes.

Figure 1. The rim is inspected using two linear phased array probes.

Inspect the wheel’s hub using a third linear phased array probe.

Figure 2. Inspect the wheel’s hub using a third linear phased array probe.

In-Service Inspection

In-service train wheels require periodic reprofiling. During this process, material is removed from the wheel to restore it to its original profile. When reprofiling is completed, the wheel rim is checked from the tread using a linear phased array probe before it goes back into service (Figure 3). Typically, a 5 MHz, 128-element phased array probe is used in this application. A linear scan is performed to cover the area of interest (Figure 4).

Checking the tread using a linear phased array probe.

Figure 3. Checking the tread using a linear phased array probe

A linear scan.

Figure 4. A linear scan.

System Performance

The inspection solution for the tread is designed to detect flaws whose size is equivalent to 1–3.2 mm diameter flatbottomed holes. The typical dead zone is 10 to 5 mm below the rim’s surface. Dead zones can be further reduced using special features in the software.

The solution enables fast wheel inspection.

Figure 5. The solution enables fast wheel inspection.

Dedicated software for wheel inspection

Figure 6. Dedicated software for wheel inspection

To meet wheel manufacturing inspection requirements, the rim can be inspected to detect flaws with a size equivalent to 3 mm diameter flat-bottomed holes with a repeatability of 2 dB, a signal-to-noise ratio (SNR) of 22 dB, and a typical dead zone of > 6 mm. The inspection speed ranges from 4 to 8 inches per second, enabling users to inspect a 36-inch wheel in less than 30 seconds. This procedure meets or exceeds international train wheel inspection standards such as EN13262, ISO5948, and AAR M-107/M-208.

Figure 8. A cabinet is used to house one or two FOCUS PX units and a computer. It includes a keyboard tray and stand and can be equipped with air conditioning, a light, and a sound alarm.

Figure 9. Evident wheel tread inspection solution, featuring a custom bubbler and immersion probe.

FOCUS PX Acquisition Unit | FocusPC™ Software & SDK

The scalable FOCUS PX acquisition unit and FocusPC software incorporate the latest phased array (PA) and conventional UT technologies for easy integration into automated and semiautomated systems. This inspection solution enables you to generate and save C-scan and A-scan raw data for applications where the verdict is based on post-inspection data analysis. This includes applications in industries such as aerospace (composite laminate), power generation (wind blade), transportation (train wheels), metal (forged parts), and more.