What is Ultrasonic Thickness gauge calibration?

Ultrasonic Thickness Gauge Tutorial

calibrating thickness gauge

An ultrasonic gauge reads the thickness of a test piece by very precisely timing echoes. To turn these time measurements into thickness measurements, the gauge must be programmed with the speed of sound in the test material and any necessary zero offset required by the instrument, transducer type, or echo shape. This process is commonly called gauge calibration. The accuracy of any ultrasonic measurement is only as good as the accuracy and care taken during calibration. Incorrect calibration will result in inaccurate thickness readings. The good news is that calibration is usually a simple process.

Many different calibrations for various materials and transducers can be stored in the gauge and quickly recalled. It is important to remember to perform a new calibration or recall an appropriately programmed calibration when the test material or transducer is changed, or if the temperature of the test material changes significantly. Also, we recommend performing periodic checks with samples of known thickness to verify that the gauge is operating properly, especially in critical applications.

Ultrasonic Thickness Gauge Calibration Chart

When using ultrasonic thickness gauges, the speed of the sound energy used to take thickness measurements depends on the test material. View the table below to see the sound velocity of different materials:

Material
Velocity (m/s)
Lead
1160–1320
Aluminum
3040–6420
Concrete
3200–3700
Wood
3300–5000
Copper
3560–3900
Brick
3600–4200
Iron
3850–5130
Glass
3950–5000
Steel
4880–5050

You can use this information to establish a pulse transit time to provide and maintain accurate thickness values.

Continue on to Velocity and Zero Calibration Thickness Gauge Tutorial Next Chapter