Remote Field Probes

Thickness and Flaw Inspection Solutions

The remote field probes include a 32 dB preamplifier. These probes feature improved wear resistance and thicker casing as well as improved differential response and SNR ratios. These probes are designed to detect irregularities such as pitting, corrosion, and erosion in ferromagnetic tubing. These probes are available in single exciter, dual pickups, and dual exciter designs.

TRS - Single exciter

Sonde_Tube_CF_TRS_01.psd Sonde_Tube_CF_TRSOS_03.psd

Features

Probe outside diameter

Probe diameters ranging from 9 mm (0.35 in.) to 26 mm (1.02 in.) are built with a protective stainless steel casing and are available in increments of 1 mm. Larger probe diameters from 22.0 mm (0.866 in.) to 50.0 mm (1.968 in.) are available in increments of 2 mm.

Probes with a diameter greater than 28.0 mm (1.100 in.) have a lightweight design with the probe body made of plastic. This design also includes stainless steel wear rings on each end and a sturdy probe cable to endure multiple probe pulls. Coils are potted (in black on picture) for maximum protection.

Probe center frequency

The most current probe central frequency is 300 (300 Hz) and it ranges from 100 Hz to 1 kHz. Other frequency ranges exist:

Probe cable

Probe cable material is manufactured in nylon (N) only.

Cable lengths are in meters: 20 m (65 ft) is the most common.

Also available in 30 m (100 ft) length.

TRX - Dual exciter

Sonde_Tube_CF_TRT_01.psd Sonde_Tube_CF_TRS_03.psd

Features

Probe outside diameter

Probe diameters ranging from 9 mm (0.35 in.) to 26 mm (1.02 in.) are built with a protective stainless steel casing and are available in increments of 1 mm. Larger probe diameters from 22.0 mm (0.866 in.) to 50.0 mm (1.968 in.) are available in increments of 2 mm.

Probes with a diameter greater than 28.0 mm (1.100 in.) have a lightweight design with the probe body made of plastic. This design also includes stainless steel wear rings on each end and a sturdy probe cable to endure multiple probe pulls. Coils are potted (in black on picture) for maximum protection.

Probe center frequency

The most current probe central frequency is 300 (300 Hz) and it ranges from 100 Hz to 1 kHz. Other frequency ranges exist:

Probe cable

Probe cable material is manufactured in nylon (N) only.

Cable lengths are in meters: 20 m (65 ft) is the most common.

Also available in 30 m (100 ft) length.

TRT - Dual pickup

Sonde_Tube_CF_CL_04.psd Sonde_Tube_CF_TRTOS_02.psd

Features

Probe outside diameter

Probe diameters ranging from 9 mm (0.35 in.) to 26 mm (1.02 in.) are built with a protective stainless steel casing and are available in increments of 1 mm. Larger probe diameters from 22.0 mm (0.866 in.) to 50.0 mm (1.968 in.) are available in increments of 2 mm.

Probes with a diameter greater than 28.0 mm (1.100 in.) have a lightweight design with the probe body made of plastic. This design also includes stainless steel wear rings on each end, and a sturdy probe cable to endure multiple probe pulls. Coils are potted (in black on picture) for maximum protection.

Probe center frequency

The most current probe central frequency is 300 (300 Hz) and it ranges from 100 Hz to 1 kHz. Other frequency ranges exist:

Probe cable

Probe cable material is manufactured in nylon (N) only.

Cable lengths are in meters: 20 m (65 ft) is the most common.

Also available in 30 m (100 ft) length.

TRC - Boiler probe

Sonde_Tube_CF_TRC_01.psd

Features

Probe outside diameter

Standard probe diameters are available from 28.0 mm (1.10 in.) to 65.0 mm (2.56 in.). The part number shows the diameter in tenths of mm (1/10 mm) so 45.0 mm = "450." Standard available diameters are listed in the chart below:

Probe diameter (catalog)
Diameter value
280
28.0 mm (1.10 in.)
340
34.0 mm (1.34 in.)
370
37.0 mm (1.46 in.)
450
45.0 mm (1.77 in.)
550
55.0 mm (2.17 in.)
650
65.0 mm (2.56 in.)

Probe center frequency

The most current probe central frequency is 300 (300 Hz) and it ranges from 100 Hz to 1 kHz. Another frequency range exists:

Probe cable

Probe cable material is manufactured in nylon (N) only.

Cable lengths are in meters: 20 m (65 ft) is the most common.

Also available in 30 m (100 ft) length.