Application Notes

Contact Us

PMI Inspection and Material Verification for Hot, In-Service Testing

Alloys are specified by design for mechanical characteristics, corrosion resistance, temperature and pressure tolerance. Positive Material Identification (PMI) can prevent mix-ups that can result in component failure, which can cause costly downtime, repair and replacement, lost (leaking) material, environmental and fire hazards, or batch contamination.

Handheld X-ray Fluorescence (HHXRF) is a fast nondestructive testing technique that measures alloy composition to confirm alloy grades. HHXRF provides rapid and definitive material confirmation when producing components, receiving alloy materials, or when installing pipes, valves, reaction vessels, or welding, bolting, and gasket joining materials.

PMI Inspection and Material Verification

Unfortunately, there is not always enough time for PMI alloy verification of mission-critical pieces when a process is down for maintenance. Consequently, there is a need for cost-effective testing techniques that can be used in-service under hot, high pressure, vibrating conditions. The DELTA Handheld XRF is the Evident Alloy PMI Analyzer of choice for these adverse PMI testing conditions.

The DELTA HHXRF is designed for high performance

  1. in high ambient temperatures;
  2. on heated, in-service components; and
  3. during heavy-duty, almost continuous testing cycles

The metal probe head and the top portion of the DELTA HHXRF act as radiators or heat sinks. The ridged aluminum surface on the top of the analyzer is thermally tied in to the primary sources of heat within the instrument. The result is an efficient channel to direct heat away from the analyzer for measuring material at temperatures up to 900 °F (480 °C).



The DELTA HHXRF’s internal failsafes, monitors, and temperature correction models ensure that analytical performance is optimized. But should any of these be overtaxed, the analyzer will not allow testing until it has cooled off.

A special, high-temperature DELTA HHXRF window is used for hot sample testing. And, for sample temperatures above 600 °F (315 °C), a tilting technique or the use of a DELTA HotHeel is recommended to minimize heat transfer even further. Consequently, a 30 seconds on – 90 seconds off duty cycle is for these situations.

Hot, in-service measurements of light elements such as silicon (Si) in alloys are particularly challenging since they can take longer to measure than heavier elements such as chromium (Cr). And, if the tilting technique or the DELTA HotHeel is used, it is even more challenging because the air path length and air attenuation have an adverse effect on light element analysis.

The DELTA series HHXRF is thermally robust enough for continuous hot, in-service PMI testing and sensitive enough to positively ID the concentrations of light elements. For silicon measurement the sensitivity of the DELTA Professional and the DELTA Premium with the new X-act Count digital pulse processor are required.

HHXRF Silicon Hot Testing Tips

Silicon (Si) is 27% of the earth’s crust. It is in soil; it is everywhere. Unfortunately, contamination from soil, dirt and debris has an adverse effect on HHXRF testing for Si. A small amount of Si on the surface of an alloy sample or on the window of the HHXRF analyzer will create an error when testing Si in low-alloy samples.

Contamination will cause a high bias in the measurement and can cause a false conclusion when concluding that Si is abundantly present when there is none or much less than the reported value. Here are some common sense tips for HHXRF Testing of Silicon.

Surface Preparation

Controlling Contamination

Best practice is to test a no silicon reference material (e.g. CP Iron) and a sample with a known small value (e.g. Carbon steel) at the beginning of each testing program and periodically (perhaps as often as every 5 samples) during testing to ensure that no contamination is present.

Final Recommendations:

To summarize: proper results require proper technique. Don’t over-tilt – less is better than more! The reference samples are the key to accurate results. Grinding and cleaning is a necessary step.

DELTA Premium and Professional Si Performance while Hot Testing (w/Hot Heel)


https://main--eds-evident-website--evident-scientific.hlx.live/en/xrf-xrd/delta-handheld/

DELTA Premium

Combining a large area, high performance Silicon Drift Detector (SDD) and a 4W advanced x-ray tube, the DELTA Premium is the ideal solution for ultra quick, analytically demanding applications such as trace levels and light elements in low alloy steel, soil, mining, and metallurgical samples.


https://main--eds-evident-website--evident-scientific.hlx.live/en/element/

DELTA Element

The DELTA Element is a cost-effective handheld X-ray fluorescence (XRF) analyzer optimized for scrap metal sorting, positive material identification (PMI), alloy QA/QC, and precious metals analysis.


https://main--eds-evident-website--evident-scientific.hlx.live/en/xrf-xrd/delta-handheld/delta-prof/

DELTA Professional

Combining a Silicon Drift Detector (SDD) and a 4W optimized X-ray tube, the DELTA Professional provides superior performance in speed, LODs, and elemental range. It analyzes Mg and higher for Alloys and Mining. The DELTA Professional is the best value solution from Evident for handheld XRF.