2 Challenges of High-Temperature Thickness Gauging and How to Solve Them

Thickness gauging high-temperature

Post Author's Photo

Sarah Williams

28 August, 2018

Crude oil is an important source of energy, but using it in its natural form is inefficient and possibly dangerous. Because of this, refineries turn crude oil into finished petroleum products.

The refining process starts with heating the crude oil and putting it into a distillation column (also known as a still). In the still, diesel, gas, and other products boil off and are recovered at different temperatures. The distillation temperature for petroleum products ranges from as low as 30 °C (85 °F) to greater than 565 °C (1,050 °F). These high surface temperatures make inspecting refinery equipment challenging.

Challenge: Too Hot to Handle

Periodic thickness measurements using ultrasonic testing (UT) are needed to monitor for corrosion and other defects, which can cause equipment failure to potentially shut down the refinery’s operation. Since the refining process runs continuously, the machinery can’t be cooled down without stopping the entire process. But high surface temperatures can easily destroy a thickness gauge transducer; many can only tolerate temperatures up to about 50 °C, or 125 °F.

Solution: Transducers That Can Take the Heat

Unlike traditional UT transducers, dual element transducers have delay material that acts as insulation between the active transducer element and the hot test surface. They’re frequently recommended for high-temperature applications because some—such as the D790—can be used on surfaces as hot as 500 °C (930 °F) with intermittent contact (5 seconds or less).

Corrosion gauging duals with temperature specifications.

Corrosion gauging duals with temperature specifications.

Challenge: Changing Sound Velocity

Accurate UT thickness measurements depend, in part, on the velocity of sound in the test material. As a material’s temperature increases, the sound velocity decreases, complicating the accuracy of the thickness readings. According to ASTM E797-95, the sound velocity of carbon steel decreases by “about 1% per 55 °C (100 °F)” increase in temperature.

To get around this problem, some inspectors will calibrate their flaw detector or thickness gauge on a hot test block. To heat the block, they’d hold it to the pipe or tank being tested. However, this process is slow, and the temperature of the block wouldn’t exactly match the test material, causing imprecise measurements.

An alternative method is to take the thickness measurements and then manually calculate a correction based on the part’s thickness and temperature. This technique is time consuming, and the operator is not able to see the true thickness reading during the inspection.

Corrosion gauging duals with temperature specifications.

Solution: Automatic Temperature Compensation

Our predecessor, Panametrics NDT, pioneered a solution to this challenge nearly 15 years ago. Today, the 38DL PLUS® ultrasonic thickness gauge includes Temperature Compensation software (as a standard option) that enables you to calibrate the gauge at room temperature and then simply input the elevated temperature of the material being tested. The gauge will automatically compensate for the velocity change and display the temperature-compensated thickness. The software is easy to use and requires little effort to set up, so you can calibrate quickly and receive real-time temperature-adjusted measurements. It’s that simple.

Post Author's Photo

Sarah Williams

Staff Writer

Sarah Williams worked for nearly a decade as a researcher and copywriter in the broadcast media industry. Now Sarah applies her skills as a writer and editor to produce compelling, high-quality material on topics related to Evident’s wide range of nondestructive testing (NDT) solutions. She writes about the latest remote visual, microscope, ultrasonic, eddy current, and phased array technologies. She also explores their applications and contributions to improving the quality and safety of the world around us. Sarah works at the office in Quebec City, where she resides with her partner, David, and her three children, Sophie, Anouk, and Éloi.