애플리케이션 노트
휴대용 XRD를 이용한 석유 및 가스 공급망 스케일링, 부식 및 슬러지 분석
XRD for Scaling, Corrosion, and Sludge Analysis
Evident provide chemists and engineers with quantitative phase analysis of scaling, corrosion, and sludge deposits onsite. Rapid characterization of scaling and corrosion products enables maintenance teams to implement the most appropriate management strategy in hours—not days. This minimizes shutdowns and can help prevent undue damage to equipment.
Scaling in Oil and Gas Assets
Scaling and corrosion affect equipment throughout the oil and gas subply chain, from exploration and extraction to pipelines and refineries. Knowing the composition of the scaling and corrosion products enables maintenance teams to apply the most appropriate anti-scaling treatment or corrosion additive immediately. For example, hydrochloric acid is often used to remove calcium carbonate scaling, while calcium sulfate scaling may be removed using chelanting agents such as ethylenediaminetetraacetic acid (EDTA).
In the past, maintenance teams would have to send samples off site for laboratory analysis, typically taking days or weeks, or would have to experiment with expensive chemical treatments (which may risk harming the equipment). However, the number of scaling and corrosion products commonly encountered within oil and gas assets is typically limited, making XRD a fast, efficient inspection option.
| Compound | Phase Name | Formation Variables |
| CaCO3 | Calcite | Partial pressure of CO2, temperature, total dissolved salts, pH |
| Aragonite | ||
| Vaterite | ||
| CaSO4.2H2O | Gypsum | Temperature, total dissolved salts, pressure, mixing incompatible fluids |
| CaSO4.1/H2O | Hemihydrate | |
| CaSO4 | Anhydrite | |
| BaSO4 | Barite | Temperature, pressure, mixing incompatible fluids |
| NaCl | Halite | Evaporation, temperature |
| FeCO3 | Siderite | Corrosion dissolved gases, pH, bacteria |
| FeS | Mackinawite | |
| FeS2 | Pyrite | |
| Fe(OH)2 | Ferrous Hydroxide |
Some common scaling and corrosion products that can be identified using portable XRD.
Sludge Deposits
Sludge deposits are often encountered in refineries and are a combination of the following:
- Hydrocarbon (e.g. lubricant and grease)
- Liquid (e.g. water and oil)
- Nonhydrocarbon or inorganic substances (e.g. scaling and corrosion)
Using XRD to understand the inorganic component of the sludge can assist the removal process and contribute to preventing reoccurrence. Dichloromethane can be used to separate the inorganic components (scaling and corrosion) from the sludge, enabling a more detailed characterization of scaling and corrosion products (by removing amorphous phases).
Benefits of Evident Portable XRD Instruments
Evident portable XRD instruments only require as little as 15 mg of corrosion/scaling products. Results are obtained in minutes and provide maintenance teams, engineers, and production chemists with the knowledge needed to select the most appropriate anti-scaling treatment or corrosion management strategy.
Benefits of TERRA® Analyzers
- Small Sample: Requires only 15 mg of sample
- Easy Sample Preparation: Does not require a skilled technician
- Fast Acquisition Time: Obtain results in only a few minutes
- Stand-Alone Instrument: No need for water cooling or a large, external power source
- No Ongoing Servicing Requirements: XRD can be performed regularly with minimal downtime
- Portability: Battery-operated, rugged design with no moving parts
Current Practice
Send Samples for Laboratory Analysis:
- 2 to 3 week turnaround
Try Different Chemical Treatment Methods:
- Using an incorrect chemical treatment may damage equipment
- Associated costs in applying the incorrect chemical treatment
- Solution in days, rather than hours
Portable XRD Solution
Send Samples for Laboratory Analysis:
- Correctly identify the scaling/corrosion products, rapidly, on site
- Select the correct treatment strategy, the first time, significantly reducing downtime
관련 제품
TERRA™ II XRD 분석기는 휴대용 정량 광물학을 제공합니다. 최대 6시간의 배터리 수명과 견고하고 비바람을 견디는 케이스를 갖춘 이 시스템은 주요 구성품과 부속 구성품을 신속하게 현장 분석하도록 설계됐습니다.