Vanta XRF 분석기를 통한 연료 중 유황 함유량의 엄격한 기준에 대한 준수

Introduction

International and domestic regulations regarding sulfur content in petroleum products and fuels have become increasingly stringent in recent years. Sulfur needs to be removed from crude oil during the refinement process since it can impact vehicles’ emission control systems, contributing to air pollution. X-ray fluorescence (XRF) analysis can be used to measure the level of sulfur in fuel. Evident’ Vanta™ XRF analyzer is well suited to meet the strict testing requirements established by standards, such as ASTM D-4294. The Vanta handheld analyzer can provide fast and accurate assessment of sulfur content in seconds and with virtually no sample preparation in even the most challenging environments.
Meeting Strict Standards for Sulfur Content in Fuel with Vanta XRF Analyzers

Using Vanta Analyzers for Low Sulfur Analysis

ASTM D-4294 establishes testing and performance guidelines when analyzing sulfur content using XRF in various petroleum products. The Vanta XRF analyzer offers several advantages for users who are testing to the D-4294 standard:

To demonstrate the Vanta analyzer’s performance capabilities, we tested several certified oil samples for low sulfur content (<150 ppm) using a VCA (Vanta C series model with silver anode). Samples were tested inside the Vanta portable Work Station for 120 seconds each. Results shown are the averages of 5 repeated tests:
Meeting Strict Standards for Sulfur Content in Fuel with Vanta XRF Analyzers
Standard
S Assay
S
S +/-
Mineral Oil
0
0
9
SMO-10-100
10
16
9
SMO-25-100
25
31,2
9
SMO-50-100
50
42,4
9
SMO-100-100
100
128,6
9
SMO-150-100
150
154
9

Conclusion

The Evident Vanta handheld XRF can effectively and accurately measure sulfur content in accordance with ASTM D-4294. The accuracy and precision provided by the Vanta analyzer helps ensure that compliance with fuel regulations is maintained.