Vanta Handheld XRF Analyzers for Industrial Lead-Based Paint Analysis*

Fast, Accurate Lead Paint Testing

For lead-based paint (LBP) testing, the Vanta™ handheld X-ray fluorescence (XRF) analyzer delivers optimum accuracy and maximum productivity. Most tests take less than 3 seconds, and even the longest tests are typically less than 10 seconds.

Because of their fast, nondestructive, and on-the-spot testing capabilities, Vanta handheld XRF analyzers rapidly screen for lead (Pb). Screen for lead with Vanta analyzers before you begin grinding, cutting, or painting. You can also use the analyzer to check your work area after clean-up.

The table below gives precision data in the context of industrial lead paint testing, where the action levels are typically in the 1.0 mg/cm2 range, and testing times are often less than 10 seconds.

The Repeatability You Need

Typical variation in results when testing a 0.3 mg/cm2 standard are excellent whether the lead is buried or on the surface. Variation decreases as test time increases.
LBP Sample Type Surface Buried (12 layers of blank paint)
3-Second Test 0.031 0.037
10-Second Test 0.017 0.027

For most paint matrices, Vanta analyzers can deliver a rapid, accurate result in 3 seconds. When identifying a challenging paint matrix through spectral analysis, the measurement time can be extended for greater accuracy and analytical confidence.

Vanta Handheld XRF for Fast, Accurate, and Nondestructive Lead-Based Paint Testing

Vanta handheld XRF lead paint analyzers enable on-site determinations of potentially dangerous levels of Pb in suspected areas. Vanta analyzers provide a better way to identify the presence of lead paint on all the usual surfaces, including wood, plaster, drywall, stucco, concrete, brick, cinder, and steel.

Get immediate feedback on the presence of lead before repair, restoration, or painting.

Tube-Based XRF Inspections of Lead-Based Paint

Compliance and the Health Hazards of Lead-Based Paint

Health hazards associated with Pb in pre-1978 structures that contain lead-based paint (LBP) are well established. Regulated limits of Pb vary from country to country and even from city to city. Laboratories, while well equipped with techniques and instrumentation to measure Pb to confirm regulatory compliance, cannot supply immediate, on-the-spot results. Laboratories can also be expensive and time-consuming.

Vanta handheld XRF analyzers dramatically reduce testing times, enabling hundreds of more tests per day on varied painted surfaces. Within seconds, Vanta analyzers can detect lead and report the results as surface area concentrations or positive or negative results based on preset values (NIST standards are available for portable XRF lead paint surface analysis).

Benefits of Vanta Handheld XRF Analyzers for Lead-Based Paint Analysis

Vanta Handheld XRF Analyzers for Industrial Lead-Based Paint Analysis*
Vanta Handheld XRF Analyzers for Industrial Lead-Based Paint Analysis*
Vanta Handheld XRF Analyzers for Industrial Lead-Based Paint Analysis*

*The Vanta handheld XRF analyzer is not HUD certified; therefore, it is not suitable for HUD LBP applications.

관련 제품


https://main--eds-evident-website--evident-scientific.aem.live/ko/vanta-for-environmental-assessments/

환경 평가용 Vanta

토양 및 퇴적물 분석용 Vanta 휴대용 XRF 분석기는 신속하고 정확한 환경 교정 및 평가 데이터를 제공합니다. 사용하기 간편한 Vanta 분석기는 가장 까다로운 현장 응용 분야에서도 견고합니다.


https://main--eds-evident-website--evident-scientific.aem.live/ko/vanta/

Vanta

Vanta™ 휴대용 XRF 분석기 시리즈는 최신의 가장 강력한 휴대용 XRF 장치로서 현장에서 실험실 품질의 결과를 요구하는 고객을 위해 신속하고 정확한 요소 분석을 제공합니다. 분석기는 IP55 또는 IP54 등급으로 견고하게 제작됐으며, 가동 시간 향상과 소유 비용 절감을 위해 낙하 시험을 거쳤습니다.