이멀젼 프로브

두께 및 결함 검사 솔루션

연락처

개요

Immersion probes are designed to be used with a water wedge or in an immersion tank when the test part is partially or wholly immersed. They are longitudinal wave probes that can be set up for refracted shear-wave inspections using a Rexolite wedge.

Advantages

  • Acoustic impedance matches water
  • Design allows fitting on water wedges for easier coupling on many surfaces and an adjustable water path (when the part to be inspected cannot be immersed in a tank).
  • Linear scanning allows coverage of 30 mm to 90 mm in one line, with very high accuracy.
  • Corrosion-resistant stainless steel case
  • Waterproof guaranteed up to 1 m (3.28 ft) under water

Typical Applications

  • Inspection of thin plate or tubing (steel, aluminum, or other)
  • Composite inspection for delamination, disbonding, etc.
  • Inline thickness gaging
  • Automated scanning

Probe Specifications and Dimensions

Part Number
Item Number
Frequency
(MHz)
Number of Elements
Pitch
(mm)
Active Aperture
(mm)
Elevation
(mm)
External Dimensions
mm (in.)
L
W
H
5L64-I1
U8330323
5.0
64
0.60
38.4
10.0
50 (1.97)
19 (0.75)
25 (0.98)
10L64-I1
U8330012
10.0
64
0.50
32.0
7.0
50 (1.97)
19 (0.75)
25 (0.98)
5L128-I2
U8330031
5.0
128
0.60
76.8
10.0
83 (3.27)
21 (0.83)
35 (1.38)
10L128-I2
U8330004
10.0
128
0.50
64.0
7.0
83 (3.27)
21 (0.83)
35 (1.38)
2.25L128-I3
U8330351
2.25
128
0.75
96.0
12.0
102 (4.02)
21 (0.83)
35 (1.38)
5L128-I3
U8330379
5.0
128
0.75
96.0
10.0
102 (4.02)
21 (0.83)
35 (1.38)
7.5L60-I4
U8330955
7.5
64
1.0
64.0
7.0
73 (2.87)
24 (0.94)
25 (0.98)

These probes come standard with an OmniScan® connector and a 2.5 m (8.2 ft) cable or can be specially fitted with other connectors and cable lengths.