COBRA 수동 스캐너

OmniScan X3 결함 탐상기와 결합한 COBRA™ 수동 스캐너는 소구경 파이프의 원주 용접 검사를 수행하는 데 사용됩니다.COBRA 소구경 스캐닝 장치는 외경이 0.84~4.5인치인 파이프를 검사하기 위해 2개의 PA 프로브를 고정할 수 있습니다.

The COBRA® manual scanner, combined with the OmniScan® PA flaw detector, is used to perform circumferential weld inspections on small-diameter pipes. The COBRA scanner holds up to two PA probes for inspections on pipes with outside diameters ranging from 0.84 in. to 4.5 in.

With its slim design, this manual scanner inspects pipes in limited ac­cess areas where minimal clearance is required. Adjacent obstructions, such as piping, supports, and structures, can be as close as 12 mm (0.5 in.). This spring-loaded scanner is designed to clasp carbon steel and stainless steel pipes of various diameters using multiple links. This unique feature enables the scanner to be installed and operated from one side of a row of pipes.The COBRA scanner is characterized by its smooth-rolling encoded movement, which enables precise data acquisition. The scanner holds up to two phased array probes for complete inspection of the weld in one pass. For pipe-to-component inspections, the scanner can be configured quickly to perform one-sided inspections using a single probe.

This Evident solution uses low-profile phased array probes with optimized elevation focusing, which enhances detection of small defects in thin-wall pipes. Specially designed low-profile wedges fitting each pipe diameter covered by the scanner are also available for a complete solution. The COBRA scanner ensures stable, constant, and strong pressure, thus providing good UT signals and precise encoding around the full circumference of the pipe.

The COBRA solution can be used with different probes, depending on the application. Phased array, TRL dual linear array, and TOFD ultrasound techniques can be performed using the following probe options:

PA Probe

The solution uses low-profile phased array probes with optimized elevation focusing, which improves the detection of small defects in thin-wall pipes.

PA Probe

The COBRA scanner can also be configured for pipe-to-component weld inspections.

PA Probe

Wedges for TOFD inpection are also available.

PA Probe

The A25 dual linear array probe (DLA) series is designed to inspect austenitic material that cannot be otherwise inspected using an A15 probe in pulse echo.

Features

FAQ

1. What is the minimum wall thickness that can be inspected with the COBRA on the smallest diameter pipe size?

We can't give an exact number for the wall thickness. It is too closely related to the defects, material type, noise in the material, and the operator's ability. The customers need to make their own tests because we cannot guarantee the performance for their specific applications. Small diameters lose a lot of power on each skip and thin wall pipe required multiple steps to cover wide welds. This combination can become a problem with very small and thin pipes. Unfortunately, we cannot give a number that will work in every situation without being too conservative.

2. Is the COBRA scanner AUT code compliant?

A definite "yes "! Scanners and instruments are one part of an overall procedure or process, which has been written to comply with the various codes. They contain features such as scan speed, encoder capability, instrument linearity, software options, and data recording that make up the equipment portion of the overall process.

COBRA helps you achieve inspection compliance for:
- ASME CC 2235
- ASME B31.3 CC 181
- ASME B31.1 CC 179

3. Can the COBRA scanner inspect pipe-to-component welds?

Yes, if the straight portion on each side of the weld is long enough, it is possible to carry out the inspection using two probes. If not, the scanner can be dismantled to perform the inspection from one side of the weld.

4. What is the straight-pipe length needed on each side of the weld to perform an inspection with the Cobra?

59 mm of straight pipe is needed for each side. From the center of the weld, the dimension needed on each side is: 59 mm + half of the weld width.