Advantages
- Design optimized for weld inspection.
- Leading edge signal to noise ratio performance.
- Wide thickness range coverage.
- New wedge concept allowing improved coupling to the part.
- Acoustically matched to Rexolite.
Typical Applications
A31 and A32 Probes
- Manual or automated inspection of 3mm to 60mm thick welds using angle beams.
- Innovative wedge design available for shear or longitudinal waves.
Probe Specifications and Dimensions
Part Number
Item Number
Frequency
(MHz)
(MHz)
Number of Elements
Pitch
(mm)
(mm)
Active Aperture (mm)
Elevation
(mm)
(mm)
External Dimensions
mm (in.)
mm (in.)
L
W
H
5L32-A31
Q3300178
5.0
32
0.60
19.2
10.0
30 (1.18)
28 (1.10)
25 (0.98)
7.5L32-A31
Q3300339
7.5
32
0.60
19.2
10
30 (1.18)
28 (1.10)
25 (0.98)
10L32-A31
Q3300340
10.0
32
0.30
9.9
10.0
30 (1.18)
28 (1.10)
25 (0.98)
2.25L32-A32
Q3300341
2.25
32
1.0
32.0
10.0
40 (1.57)
28 (1.10)
26 (1.02)
5L32-A32
Q3300180
5.0
32
1.0
32.0
10.0
40 (1.57)
28 (1.10)
26 (1.02)
5L64-A32
Q3300179
5.0
64
0.50
32.0
10.0
40 (1.57)
28 (1.10)
26 (1.02)
Colspan=10
These probes come standard with an OmniScan® connector and a 2.5 m (8.2 ft) cable or can be specially fitted with other connectors and cable lengths.