Portable XRD and XRF for Cement Analysis

By Josh Litofsky

Evident portable X-ray diffraction (pXRD) and X-ray fluorescence (pXRF) analyzers provide technicians, engineers, and scientists with real-time, quantitative identification for a wide range of cement, concrete, and individual components. With onsite analysis, these compact instruments can identify and quantify the material phases quickly and accurately.

While sending samples to an outside lab can take days or weeks, cement analysis using Evident’s patented pXRD and pXRF technology takes minutes or seconds. Using these instruments for cement production and quality control can help you identify the value of material feeds from quarries, stockpiles, and silos to the kilns and the mills using blending processes.

Quarry

Raw Materials Storage

Corrective Materials

Gypsum

Clinker Storage

Kiln Relining

Benefits of Evident XRD

Evident pXRD technology has many benefits, including:

Evident XRD analyzers provide similar analytical capability as larger, more complex lab systems in compact, easy-to-use, cost-effective solutions.

Quantification in the Field

Evident pXRD analyzers can identify all mineral phases in the field in real-time, regardless of chemistry and orientation. A variety of materials can be used with our Cu-tube X-ray diffraction analyzers, including Portland cement, concrete, gypsum, limestone, pozzolan, and more:

Evident XRD analyzers can test a variety of materials for the cement production process.

Engineers can use this information on-site in real-time to better understand the chemical ratios and grades for cement, along with analysis of any additives used during the process. At the plant, engineers can obtain a chemistry or phase analysis on location and reduce the amount of out-of-spec clinker, raw material, and additives that need to be held and re-blended.

As shown in the diffractogram above, the Evident II and III XRD analyzers can quickly identify a variety of phases in the cement and aggregate industry. This includes calcium levels at the quarry, sulfate content for the gypsum feed stream, and any additives or imperfections that might be encountered along the way. Each component can be identified and quantified in real-time using Evident’s innovative and integrated automated phase ID and quantitative software, or with standalone software packages such as XPowder (included). These options make it even easier and faster to quantity the phase information and minerology during the entire process.

A technician quickly runs a test using the TERRA II portable XRD analyzer.

Additional Phase Identification

Concrete and cement are mostly composed of calcium silicates, sulfates, and carbonates, along with silica. While Evident pXRD analyzers can easily and quickly quantify these components, they can also identify other phases in the product. These phases include sulphospurrite, also referred to as ternesite.

High levels of sulphospurrite lead to increased energy costs and decreased efficiency during production. Our TERRA II and BTX III pXRD analyzers can quickly and easily identify and quantify sulphospurrite, as shown below.

XRD analysis of Portland cement identifies and quantifies sulphospurrite.

In addition to the desired calcium silicate phases of alite, larnite, and hatrurite, there is a sizeable (20%) concentration of sulphospurrite contaminate in commercial-grade Portland cement. This, and similar intermediate phases such as alkali chlorides or alkali sulphates, will occur due to temperature differentials in the kiln or cyclone during production.

When these sulphospurrite rings are found in the fouling of kilns, they can cause various problems, such as damaging the kiln seal, increasing energy consumption, allowing more dust to get into the kiln, and decreasing safety. Over time, these fouled kilns may lead to an expensive unplanned downtime, so it is important to identify these potentially contaminate phases early.

The Bogue Calculation

The Bogue calculation is used to estimate the four main minerals in the clinker quickly and easily in Portland cement: alite, belite, aluminate, and ferrite based on concentrations of various oxides, including silicon oxide, iron oxide, aluminum oxide, and calcium oxide. The Bogue calculation is shown below:

Alite (Ca3Si): C3S = 4.0719CaO - 7.6024SiO2 -1.4297Fe2O3-6.7187Al2O3

Belite(Ca2Si): C2S = 8.6024SiO2 + 1.0785Fe2O3 + 5.0683Al2O3 - 3.0710CaO

Aluminate (Ca3Al): C3A = 2.6504Al2O3 - 1.6920Fe2O3

Ferrite (Ca4AlFe): C4AF = 3.0432Fe2O3

Evident Vanta™ portable XRF analyzers can quantify these compositions in 30 seconds or less of run time. Using the commercial Portland cement sample, you can calculate the unknown concentrations quickly and directly on the unit.

Vanta pXRF analyzers can help you easily test for critical components of clinker in cement in real-time without any external supplies, as shown above. Combined with the TERRA II or BTX III pXRD analyzers, these instruments provide you with a complete tool kit for concrete and cement analysis.

Cement analysis is quick and easy using a Vanta XRF analyzer.

Vanta XRF analyzers are rugged, reliable instruments for the cement manufacturing process.

Learn More about Portable XRF and XRD

Want to learn more about pXRF and pXRD in the cement production process? Reach out to your local sales representative to set up a demo or check out the following resources:

Evident, the Evident logo, Vanta, BTX, and TERRA are trademarks of Evident Corporation or its subsidiaries.

SwiftMin is a registered trademark of the MinEx CRC.

Post Author's Photo
https://main--eds-evident-website--evident-scientific.aem.live/de/authors/josh-litofsky/

Josh Litofsky

Application Scientist, XRF Technologies

Josh Litofsky holds a bachelor’s degree in physics from Beloit College and PhD in chemical engineering from Pennsylvania State University. For his PhD, he focused his research on advanced characterization of designer materials using X-ray diffraction. From 2019 to 2022, Josh brought his expertise to Evident as an application scientist, supporting our X-ray fluorescence (XRF) analyzers to provide enhanced solutions to customers. In his free time, Josh enjoys running and has run the fastest 100k in the state of Pennsylvania.

Verwendete Produkte

Vanta

Die Handanalysatoren für die Röntgenfluoreszenzanalyse der Vanta Serien sind unsere neusten und leistungsstärksten RFA-Handanalysatoren. Sie liefern eine schnelle und genaue Analyse chemischer Elemente für Kunden, die Ergebnisse von Laborqualität im Außeneinsatz benötigen. Die Analysatoren verfügen über ein robustes Design, das für die Schutzart IP55 oder IP54 ausgelegt ist, und sie sind Falltest getestet für längere Betriebszeiten und niedrigere Betriebskosten.