Application Notes

Contact Us

Measuring the Wall Thickness of Small-Diameter Metal and Plastic Tubing

Measuring the Wall Thickness of Small-Diameter Metal and Plastic Tubing

This application note explains how to obtain the wall thickness and concentricity of small-diameter metal and plastic tubing. Learn about the ultrasonic gauging equipment and procedures used to take small-diameter tubing measurements.

Measuring Wall Thickness for Small Metal and Plastic Tubing

Small metal and plastic tubing with a diameter less than 3 mm or 0.125 in. is used in a wide variety of industrial and medical applications. Because the tube walls are so thin, manufacturing tolerances for thickness and concentricity are typically very tight. The correct wall thickness is especially important for safety-critical products, as well as tubing used in medical environments. Ultrasonic gauging offers a fast and nondestructive alternative to the common process of cutting tubing samples for optical measurement of thickness.

Equipment Used for Small-Diameter Tubing Measurements

The recommended instrument for small-diameter tubing measurements is the 72DL PLUS™ or 39DL PLUS™ ultrasonic thickness gauge. The 72DL PLUS gauge offers higher precision and higher frequency probes for thinner wall thickness, such as single or multilayer diameter tubing with wall thickness down to 0.001 in. (25 μm). To efficiently couple sound energy into small-diameter test pieces, inspectors must focus the sound beam into a very small spot. When diameters are under approximately 3 mm (0.125 in.), this process is usually done with a 20 MHz focused immersion transducer that focuses the sound beam through a column of water. The transducer is commonly held in a V-notch bubbler that positions the tube sample in the center of the beam. The RBS-1 recirculating bubbler system is a convenient tank, pump, and bubbler mounting fixture for benchtop immersion measurements.

Ultrasonic Gauging Procedure for Small Metal and Plastic Tubing

The optimal transducer and instrument setup for a given tubing product are selected after initial testing of product samples.

Using a transducer such as a M316-SU (20 MHz, F=.75 in.) and an appropriate fixturing device such as a bubbler, the 39DL PLUS gauge is set up for a Mode 3 (metal) or Mode 2 (plastic) measurement. This setup involves adjustment of gain and blanking parameters for optimal echo detection. The minimum measurable wall thickness is typically 0.1 mm (0.004 in.) for diameters as small as 1.5 mm (0.060 in.).

For thinner thickness measurements, use the 72DL PLUS gauge. If a V-notch bubbler is used, the tubing can be rotated to quickly obtain thickness measurements all around the circumference as a concentricity check and slid to measure thickness at different points along the length. Since this process is nondestructive, there is no need to cut long sections of tubing for measurement.

The examples below show typical measurements of small metal and plastic tubing with a 39DL PLUS gauge.

Steel tubing, 1.5 mm (0.060 in.) diameter, 0.15 mm (0.006 in.) wall thickness

Steel tubing: 1.5 mm (0.060 in.) diameter, 0.15 mm (0.006 in.) wall thickness

screen readout - plastic tubing

Plastic tubing: 2 mm (0.080 in.) diameter, 0.37 mm (0.014 in.) wall thickness


https://main--eds-evident-website--evident-scientific.aem.live/en/39dl-plus/

39DL PLUS

The 39DL PLUS™ gauge is field-proven and reliable for virtually every ultrasonic thickness inspection. It works as a precision or corrosion thickness gauge to meet the challenges of demanding applications across industries, including maintenance, manufacturing, and energy production. With world-class ultrasonic capability, fast scanning speeds, and fully integrated wireless connectivity, this handheld gauge delivers results you can trust in a seamless workflow.


https://main--eds-evident-website--evident-scientific.aem.live/en/72dl-plus/

72DL PLUS

The 72DL PLUS™ advanced ultrasonic thickness gauge delivers precision thickness measurements at high speed in a portable, easy-to-use device. Compatible with single element transducers up to 125 MHz, this innovative instrument is ideally suited to measure the thickness of ultra-thin materials, including multilayer paint, coatings, and plastic. It can simultaneously display the thickness of up to 6 layers.


https://main--eds-evident-website--evident-scientific.aem.live/en/45mg/

45MG

The handheld 45MG ultrasonic thickness gauge is packed with measurement features and software options. This unique instrument is compatible with the complete range of Evident dual element and single element transducers, making this gauge an all-in-one solution for virtually every thickness gauge application.