애플리케이션 노트
항공기 폴리머 창 두께 측정
Background: Multilayer plastic construction is commonly employed by the aerospace industry for aircraft windshields, in which two relatively thick layers of acrylic or a similar hard plastic are typically sandwiched around a thinner layer of a soft material such as polyvinyl butyral (PVB) which provides thermal stress relief. The hard plastic layers are typically about 0.4 in. to 0.5 in. (10 to 12.5 mm) thick and the soft inner layer about 0.050 in. (1.25 mm) thick.
Scratches in aircraft windshields are typically removed by polishing during routine maintenance, after which thickness measurement is required to insure that windshield thickness has not been reduced below a specified minimum. Additionally, manufacturers need to measure the soft PVB middle layer to insure that it is within specified thickness.
Equipment: The Model 38 DL PLUS thickness gage with the Multi-Measurement software option is recommended for measurement of individual layers and total part thickness in this application. With this software, the 38DL PLUS gage is capable of using separately programmed setups (including sound velocity, gain, and blanking settings) for each layer being measured to optimize performance. In this test, the gage is normally used with either an M110-RM (5 MHz) or M112-RM (10 MHz) contact transducer.
Typical Procedure: The waveforms below show simultaneous measurement of all three layers of a typical multilayer aircraft windshield using an M110 (5 MHz) contact transducer. The outside layer is measured in Mode 1, to the first negative echo. The thin middle layer is measured in Mode 2, negative peak to positive peak. The bottom layer is also measured in Mode 2, positive to negative. The small echo just to the left of the detected backwall echo in the bottom screen print is not another layer, but rather a second multiple of the backwall echo from the outer layer.
outside polycarbonate layer
middle PVB layer
bottom polycarbonate layer
As with any ultrasonic thickness measurement, accuracy is dependent on proper sound velocity calibration. Velocity calibration must be performed for each material being measured, on samples of known thickness.
관련 제품
https://main--eds-evident-website--evident-scientific.aem.live/ko/72dl-plus/
72DL PLUS
72DL PLUS™ 고급 초음파 두께 측정기는 사용이 간편한 휴대용 장치로 빠르고 정밀하게 두께를 측정합니다.최대 125MHz의 단일 요소 탐촉자와 호환되는 이 혁신적인 두께 측정 도구는 다층 도장, 코팅, 플라스틱과 같은 초박막 소재의 두께의 측정에 매우 적합합니다.최대 6층의 두께를 동시에 표시할 수 있습니다.
https://main--eds-evident-website--evident-scientific.aem.live/ko/45mg/
45MG
45MG 고급 초음파 두께 측정기에는 표준 측정 기능은 물론 다양한 소프트웨어 옵션이 있습니다.이 독특한 두께 측정 도구는 당사의 이중 요소 및 단일 요소 두께 측정 탐촉자와 호환됩니다.
https://main--eds-evident-website--evident-scientific.aem.live/ko/38dl-plus/
38DL PLUS
다용도로 활용 가능한 38DL PLUS™ 측정기를 이중 요소 탐촉자와 함께 사용하면 부식된 파이프의 두께를 측정할 수 있으며, 단일 요소 탐촉자를 사용하면 박층 또는 다층 소재의 두께를 매우 정확하게 측정할 수 있습니다.