3 Tips for Accurate Small-Diameter Tubing Measurements
Small-diameter metal and plastic tubing is used in a wide variety of industrial and medical applications. Some of these tubes can be smaller than 0.125 inches (3 mm) in diameter, with very thin walls. When their applications are safety-critical, they need to meet tight manufacturing tolerances for wall thickness and concentricity. In the past, manufacturers would have to cut the samples to visually inspect the tube or measure the wall thickness using calipers. Today, there’s a fast and nondestructive alternative for quality control inspections of small-diameter tubing, ultrasonic (UT) thickness gaging.
Recommended UT Thickness Gage, Equipment, and Inspection Setup

B103 V-notch bubbler with the RBS-1-15 recirculating bubbler system
We recommend using an Evident 38DL PLUS® or 45MG ultrasonic thickness gage (using the 45MG-SE single element and 45MG-WF waveform software options) along with a 20 MHz M316-SU immersion transducer with a 0.75 in. (19 mm) focus for this application.
Small-diameter tubing inspections can be easy when using the proper equipment and correctly calibrated instruments, but measurement errors may still occur. Here are 3 simple tips to help you avoid common issues.
- Use the IF Blank function to avoid the interface echo trailing edge

The trailing edge of the large interface echo interfering with the first back wall echo
The IF Blank setting can be adjusted to select the strongest pair of signals. Although the selected pair of signals changes as the IF Blank is adjusted, the time of flight between successive back wall echoes remains the same.



Varying IF Blank settings results in the same thickness measurement
- Watch for varying back wall signal amplitude when using a focused immersion transducer

3rd and 4th multiples are the highest amplitude due to the focal length
- Don’t set the M3 Blank out too far

Detection between the 3rd and 5th back wall echoes

Detection between the 3rd and 4th back wall echoes