Ultrasonic Thickness Measurements of Hot Metals

Ultrasonic thickness gauging is not just limited to materials at normal (ambient) temperatures. The thickness of a part can be measured even when the surface temperature approaches 900 deg F (500 deg C). In metals particularly there may be circumstances when the thickness of the part needs to be measured during an ongoing process where the test piece cannot be cooled down.

Ultrasonic testing has the advantage - just as at normal temperatures - that thickness measurements can be taken from one side of the hot part. Ultrasonics may be the preferred method of measurement even when both sides of the hot metal part are accessible. The reason is that ultrasonic thickness measurements can be made instantly and with very brief contact with the hot surface.

For the best calibrated accuracy, the effect of the higher temperature on the velocity of sound must be taken into account. Although change in sound velocity with temperature is much less in metals than in plastics, significant errors in measured thickness can result if sound velocity is not corrected.

Equipment: A major reason to classify hot metals separately is that when the temperature of a surface exceeds approximately 120 deg F (50 deg C), Microscan contact transducers cannot be used because they will be damaged by thermal expansion effects. Consequently, gauges that exclusively utilize contact transducers cannot be used for making thickness measurements on hot surfaces.

For measurement of smooth hot metal we recommend the Model 38DL PLUS, or Model 45MG with Single Element software, which can be used with delay line transducers or immersion transducers. On hot surfaces delay line transducers must be used with selected high temperature delay lines. The thickness of hot corroded metal parts or parts with rough surfaces should be measured with a D790 or D791 dual transducer. The 38DL PLUS and , 45MG gages can both be used for this purpose.

相关产品


https://main--eds-evident-website--evident-scientific.aem.live/zh/45mg/

45MG

45MG是一款配备有标准测量功能和多个软件选项的高级超声测厚仪。这款独特的厚度测量工具可与我们的全系列双晶和单晶测厚仪探头兼容。


https://main--eds-evident-website--evident-scientific.aem.live/zh/38dl-plus/

38DL PLUS

用途广泛的38DL PLUS测厚仪既可与双晶探头一起使用,测量被腐蚀管道的厚度,也可与单晶探头一起使用,对薄材料或多层材料进行非常精确的厚度测量。


https://main--eds-evident-website--evident-scientific.aem.live/zh/27mg/

27MG

27MG是一款初级超声波测厚仪,设计用于从内部受到腐蚀或侵蚀的金属管道或工件的一侧进行准确的厚度测量。这款测厚仪轻巧耐用,设计符合人体工程学的要求,使用单手就可以方便地进行操作。