Fuel Oil Analysis Using Handheld XRF: An Experiment

february-2020-dillon-mcdowell-banner.jpg

Post Author's Photo

Dillon McDowell

11 February, 2020

Fuel Oil Requires Careful Monitoring

On-site testing of fuel oil has become an important task for machine and engine operators. Globally, stricter air pollution regulations require careful monitoring of sulfur content in fuels. In addition, the analysis of wear metals and fuel additives helps operators avoid costly downtime caused by engine failures.

Energy-dispersive X-ray fluorescence (EDXRF), typically referred to as X-ray fluorescence or XRF, is a common technique used for these tasks. In simple terms, XRF is a nondestructive technique to measure the elemental composition in a sample. XRF was traditionally performed on benchtop systems, but today handheld XRF analyzers like our Vanta™ series are commonly used.

Vanta handheld XRF analyzers provide rapid elemental analysis and alloy identification.

The Experiment: Fuel Oil Analysis Using Handheld XRF Analyzers

To gauge how effective Vanta handheld XRF analyzers are at identifying various elements in oil samples, we ran an experiment. We used the analyzer to test certified reference material (CRMs) samples for oil screening (mineral-oil-based). Two sets of samples were prepared in a 100 ml plastic bottle with a Prolene window cap for analysis:

The Results

Our results demonstrated that handheld XRF can appropriately measure sulfur at low levels in accordance with ASTM D-4294 precision requirements. We also demonstrated that it can detect various wear metals and additive elements at low ppm levels—a performance that is comparable to several benchtop systems.

Here are some other key insights:

Wear Metals and Additives

Sulfur Analysis

5 Important Takeaways About Fuel Oil Analysis Using Handheld XRF

To wrap up, here are five important conclusions:

  1. Overall, handheld XRF can rapidly screen for multiple elements of interest in both fuel and lubricating oils.
  2. Interferences need to be considered during analysis, as they may bias results. Follow method recommendations for bias and drift correction.
  3. Sample preparation is key. Improper sample handling can have a severe impact on results, particularly for lighter elements like sulfur (D4294 also warns about this).
  4. Handheld XRF is ultimately a screening tool. Some lab testing will still be merited where possible.
  5. As a best practice, always consider your testing goals and tolerance for risk when doing any regulatory-driven testing.

Make sure to subscribe to the InSight Blog (if you haven’t already!) to stay informed about the latest XRF and NDT technology.

Post Author's Photo

Dillon McDowell

Senior Applications Scientist, Analytical Instruments

Dillon McDowell is a senior applications scientist at Evident, specializing in using X-ray fluorescence (XRF) and other analytical technologies in alloy, precious metals, and regulatory (RoHS and consumer product safety) applications. Before joining Evident, Dillon was a research assistant at Northeastern University’s Nanomagnetism Research Group. Dillon has a Bachelor of Science in physics and a Master of Science in mechanical engineering from Northeastern University and has been published in the Journal of Material Chemistry C.