Inspection of Tee-Joint Welds on Bridge Members

General Description of the Application

During the manufacture of bridge structures, a lot of automated welding is used to rapidly and efficiently build multiple parts. One of these welding processes is a system that uses opposing arcs to weld stiffeners to webs on the bridge parts. This welding process produces horizontal fillet welds simultaneously on both sides of a tee-joint.
One problem with this technique is that on thin stiffeners, the heat exchange on vertical parts causes the parts to crack in the upper part of the weld.

The best method to inspect this particular defect without doing false calls on non-rejectable defects is to use an ultrasonic phased-array technique. The multiple-element probe can scan from 40 to 70 degrees simultaneously, allowing a user to inspect the complete volume of the weld in a single pass. When combined with the portable phased-array unit and the appropriate probe, the technique offers fast inspection and exact positioning of defects. Compared to conventional UT, this technique enables any inspector, with a few hours of training, to operate at least five times faster and present more reliable results.

Typical Inspection Requirements

• Thickness of the stiffeners: 10 mm (0.375 in.) to 16 mm (0.629 in.) (or more)
• Material: steel
• 38.1 mm (1.5 in.) of free space over the weld on the stiffeners
• Real-time A-scan and sectorial scan imaging
• User-friendly operating system

Defects Detected

• Crack
• Porosity
• Lack of fusion

Description of the Solution

• Manual inspection using one small phased-array probe
• One-line scan at around 25 mm/s (1 in.), one side at a time
• Inspection with 40- to 70-degree refracted angle
• Real-time display of sectorial scan and A-scan

Material Requirements

• Any //OmniScan SX or phased array instrument
• 5-MHz, 16-element probe
• Rexolite® wedge
• Couplant

Inspection Method

The setup is configured to perform an azimuthal scan from 40 to 70 degrees. The display and gain level are adjusted using a representable calibration standard.

The scan then performs one longitudinal scan about 2 mm (0.078 in.) over the weld on each side of the stiffeners. Data is displayed in real time and can be logged into a report sheet or marked directly on the part for immediate repair.

Note

Due to the multiple angles and the focusing capabilities, this technique is not based on amplitude detection, but on signal recognition of each type of defect, added to an accurate positioning of the defect in the weld.

Client´s Comments

With this method, we have been able to improve the efficiency, the precision, and the quality of our inspections. Furthermore, this new technique allows the technician to evaluate defects more easily. Now, the phased-array ultrasound technology is helping us evolve as ultrasonography did for medicine.

相关产品


https://main--eds-evident-website--evident-scientific.aem.live/zh/phasedarray/omniscan-x3/

Omniscan X3 Series

OmniScan X3系列的每台探伤仪都是一款功能齐备的相控阵工具箱。其创新型全聚焦方式(TFM)和高级相控阵(PA)功能助力您充满信心地识别缺陷,其性能强大的软件功能和便捷的工作流程有助于提高您的检测效率。


https://main--eds-evident-website--evident-scientific.aem.live/zh/omniscan-sx/

Ominiscan SX

轻便的单组Omniscan SX探伤仪装有一个方便用户阅读的8.4英寸(21.3厘米)触摸屏,可提供性价比很高的检测解决方案。OmniScan SX有两种型号:SX PA和SX UT。SX PA是一个16:64PR仪器,它与仅使用UT技术的SX UT一样,配备有一个用于脉冲回波、一发一收或TOFD(衍射时差)检测的常规UT通道。